Signal processing for an infrared array detector

Young, M. A. ; Smith, G. E. ; Pimentel, G. C.

[S.l.] : American Institute of Physics (AIP)
Published 1989
ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
Electrical Engineering, Measurement and Control Technology
Notes:
A broadband detection scheme for a time-resolved infrared absorption spectrometer, based on a multielement mercury-cadmium-telluride (MCT) array, has been successfully implemented. The spectrometer achieves a resolution on the 10-ns time scale despite the much larger time constant characteristic of the MCT elements. Our signal-collection circuitry takes advantage of the slow decay by integrating the detector response to pulsed IR radiation. The dynamic range is 100–1 and the resultant noise level is near the detector limit. Data acquisition for the 120 elements is fast enough to allow scan rates of 30–40 Hz. The completed electronics are sufficiently compact to be situated local to the array detector, and the design is relatively inexpensive to construct using commonly found electronic components.