A high counting rate parallel-plate gas proportional counter for fluorescence extended x-ray-absorption fine structure studies : Proceedings of the 4th international conference on synchrotron radiation instrumentation

Smith, A. D. ; Padmore, H. A. ; Buksh, P. A.

[S.l.] : American Institute of Physics (AIP)
Published 1992
ISSN:
1089-7623
Source:
AIP Digital Archive
Topics:
Physics
Electrical Engineering, Measurement and Control Technology
Notes:
Gas proportional counters (GPCs) are well suited for fluorescence extended x-ray-absorption fine structure (EXAFS) experiments in the soft x-ray (〈4 keV) region. They have a near unity detection efficiency and offer good energy resolution. Of the different types of GPC, the parallel-plate device is an attractive option for use with synchrotron radiation beam lines where its intrinsically high count rate permits optimum use of the high photon flux available. The development of a parallel-plate gas proportional counter for use on soft x-ray beam lines at the Daresbury synchrotron source is described. This device is demonstrated to have a counting rate in excess of 1 MHz, and data are presented on its performance including its energy resolution, long-term gain stability, and sample EXAFS spectra.
Type of Medium:
Electronic Resource
URL: