Development of fast response calorimeter for neutral beam shine-through measurement on CHS : Papers from the thirteenth topical conference on high temperature plasma diagnostics
Osakabe, M. ; Takeiri, Y. ; Takanashi, T. ; Tsumori, K. ; Okamura, S. ; Matsuoka, K. ; Akiyama, R. ; Asano, E. ; Kaneko, O. ; Kawamoto, Y. ; Oka, T.
[S.l.] : American Institute of Physics (AIP)
Published 2001
[S.l.] : American Institute of Physics (AIP)
Published 2001
ISSN: |
1089-7623
|
---|---|
Source: |
AIP Digital Archive
|
Topics: |
Physics
Electrical Engineering, Measurement and Control Technology
|
Notes: |
A fast response calorimeter has been developed for the neutral-beam (NB) shine-through measurement. This calorimeter has the following advantages. (1) Temporal variation of the heat load onto the calorimeter can be measured. (2) Measurement under the relatively high heat flux environment is possible. (3) The calorimetric measurement under continuous and steady-state heat load environment is also possible. The verification of the measurement principle was done using an NB-injection system on the compact helical system (CHS) at the National Institute for Fusion Science (NIFS). The measured NB power densities are compared to the power densities being evaluated by the CHS-NB profile database. It was experimentally confirmed that the time constant of the measurement is about 7 ms. © 2001 American Institute of Physics.
|
Type of Medium: |
Electronic Resource
|
URL: |