Effects of hydrocarbon quenching in gas chromatography when using the flame photometric detector
ISSN: |
1612-1112
|
---|---|
Keywords: |
Gas chromatography ; Flame photometric detector ; Sulfur compounds ; Response quenching
|
Source: |
Springer Online Journal Archives 1860-2000
|
Topics: |
Chemistry and Pharmacology
|
Notes: |
Summary The long time retardation of the main hydrocarbon peak in the chimney of the flame photometric detector greatly reduces the responses of later-eluting sulfur compounds. In the absence of hydrocarbons in the flame, the slope (s) of the log I vs. log [S] plot (where I is the sulfur response and [S] is the sulfur concentration in the sample) is of the highest value and is constant for all experimental conditions tested. Flame hydrocarbons cause the s value to decrease, and this is dependent on the oxygen to hydrogen ratio in the flame (O/H) and, under certain conditions, also on the sulfur to carbon ratio (S/C) of the sample. The abnormalities observed in the determination of methyl thiol in natural gas are explained on the basis of the present study.
|
Type of Medium: |
Electronic Resource
|
URL: |