Rapid SPME/LC/MS/MS Analysis of N-Methylcarbamate Pesticides in Water

Volmer, D. A. ; Hui, J. P. M.
Springer
Published 1998
ISSN:
1432-0703
Source:
Springer Online Journal Archives 1860-2000
Topics:
Energy, Environment Protection, Nuclear Power Engineering
Medicine
Notes:
Abstract. Solid-phase microextraction combined with fast short-column liquid chromatography/electrospray ionization tandem mass spectrometry (SPME/LC/MS/MS) was used for isolating and analyzing nine N-methylcarbamate pesticides from water samples. Several SPME parameters, such as polarity of fibers, extraction time, and effect of ionic strength, were investigated and their impact on the SPME/LC/MS/MS technique was studied. The method was shown to be sensitive with detection limits between 0.3 and 1.9 μg/L and reproducible with precision between 4.5 and 12.7% RSD. The versatility of the method was exhibited by its excellent linearity in the concentration range of 2–2,000 μg/L in drinking water. A comparison of the SPME/LC/MS/MS method with LC/MS/MS methods utilizing traditional sample preparation techniques shows that the former offers similar performance in terms of precision and linearity, but is clearly easier to use and faster to perform.
Type of Medium:
Electronic Resource
URL: