Correlation of retention times on liquid crystal capillary column with reported vapor pressures and half-lives of compounds used in pheromone formulations

Heath, R. R. ; Tumlinson, J. H.
Springer
Published 1986
ISSN:
1573-1561
Keywords:
Insect pheromone formulations ; liquid crystal capillary columns ; half-lives of compounds ; vapor pressure
Source:
Springer Online Journal Archives 1860-2000
Topics:
Biology
Chemistry and Pharmacology
Notes:
Abstract A method has been developed to determine by capillary gas chromatography on liquid crystal stationary phases the relative vapor pressures and half-lives of many compounds used as insect pheromones. This study demonstrated that the retention time of seven acetates on a liquid crystal column (cholesteryl-p-chlorocinnamate) could be correlated closely to the reported vapor pressures of the compounds. For 13 additional pheromonal acetates and alcohols, reported half-lives showed a high degree of correlation with their retention times on the liquid crystal column. Thus chromatography on capillary liquid crystal gas Chromatographie columns appears to be a useful method for determining the relative volatilities of many pheromones to facilitate the development of more precise formulations.