Self-tonometry: technical aspects of calibration and clinical application

Groenhoff, S. ; Draeger, J. ; Deutsch, C. ; Wiezorrek, R. ; Hock, B.
Springer
Published 1992
ISSN:
1573-2630
Keywords:
error detection system ; optical sensor ; self-tonometer ; self-tonometry
Source:
Springer Online Journal Archives 1860-2000
Topics:
Medicine
Notes:
Abstract Self-tonometry could be established as an important diagnostic tool in the early diagnosis and follow-up of glaucoma provided that a self-tonometer is precise, safe and easy to use. Furthermore, it has to comply with the legal standards for tonometer calibration. A new software-controlled detection system for the self tonometer permits the evaluation of external factors during the measurement which influence the precision of the tonometric readings. The updated version of the self-tonometer (ST) was compared to the Draeger hand-held applanation tonometer (HAT) in a clinical study with 82 patients (151 eyes). The results show a regression line characterized by a slope of 0.99, an y interception of 1.41, a correlation coefficient of 0.96 and a standard deviation for the ST readings of ±1.78 mmHg. Self-tonometry improves the IOP monitoring as well as patient compliance and therefore is a potential diagnostic tool in the management of glaucoma.
Type of Medium:
Electronic Resource
URL: